Anais de Pesquisa Clínica e Laboratorial

  • ISSN: 2386-5180
  • Índice h do diário: 17
  • Pontuação de citação de diário: 6.26
  • Fator de impacto do periódico: 5.31
Indexado em
  • Genamics JournalSeek
  • Infraestrutura Nacional de Conhecimento da China (CNKI)
  • CiteFactor
  • Diretório de Indexação de Periódicos de Pesquisa (DRJI)
  • publons
  • Euro Pub
  • Google Scholar
  • SHERPA ROMEU
  • Laboratórios secretos do mecanismo de pesquisa
Compartilhe esta página

Abstrato

Evaluation of Point of Care Test (POCT), i-CHROMA™ Serum C-Reactive Protein (CRP) Assay and Microalbumin Urine (MAU) Methods

Suman Bains, Chidi Anyaeche, Angus Wyatt, Olu Coker and John Bolodeoku

Objectives: The objective of this study was to evaluate the performance of the Boditech i-CHROMA™ point-of-care testing (POCT) method for the quantification of serum CReactive Protein (CRP) and Urine Microalbumin (MAU) compared well with the traditional Abbott Architect Ci8200 laboratory method.

Design and Methods: Serum samples of CRP (n=44) and urine samples for MAU estimation (n=25) were analysed at the Homerton University Hospital using the routine laboratory method (Abbott Architect Ci8200) and the Boditech i-CHROMA™ POCT method.

Results: The i-CHROMA™ CRP and MAU method showed good correlation with that of the Abbott Architect Ci8200: r2=0.905 and r2=0.987 respectively. Overall, there was a slight negative bias seen with the i-CHROMA™ CRP method (bias=-8.15, bias difference=-9.05%) and a negative bias seen with the i-CHROMA™ MAU method (bias=35.84, bias difference=-27.88%).

Conclusion: The Boditech i-CHROMA™ POCT method provide a reliable measurement of CRP and MAU and compared well with the traditional Abbott Architect Ci8200 laboratory method.

Isenção de responsabilidade: Este resumo foi traduzido usando ferramentas de inteligência artificial e ainda não foi revisado ou verificado